Andre Minoli← All projects
Hemosonics Quantra · Medical technology

Clarity when every decision counts.

Designing a more understandable interface and a simpler sample-loading experience for a surgical hemostasis analyzer.

RoleLead designer
ScopeUX/UI · Industrial design · Usability
TeamKey Tech & Hemosonics
Quantra hemostasis analyzer
Quantra hemostasis analyzer · Interface and industrial design
The challenge

Two moments. One need for confidence.

A working device was only the beginning. We needed to understand how people loaded a sample and interpreted the results under pressure.

01 · SAMPLE LOADING

Make the physical task simpler.

Workflow observation revealed a process that was complex, time-consuming, and vulnerable to error.

02 · DATA PRESENTATION

Make critical information legible.

Results required a steep learning curve. The interface needed to support understanding at a glance.

Observed workflow from preparing and loading a sample through reading results
Mapping the workflow helped identify time-consuming, stressful, and error-prone steps.
01 · Interpreting results

Make the important state easier to see.

I explored ways to communicate wide-ranging results at a glance, looking to flight controls for inspiration and refining dial concepts through user feedback.

The final direction emphasized the active range and critical information. Color was supported by scale, shape, and an icon so meaning did not depend on color alone.

Interactive 1:1 prototypes let testing locations review updated versions remotely as the interface evolved.

Quantra interface with dials showing clotting results and highlighted states
Interface direction · The active range and out-of-range state are visually emphasized.
02 · Loading a sample

A simpler workflow introduced a new physical challenge.

The team explored replacing a syringe connection with a vacutainer. The workflow could be simpler, but piercing the tube required approximately 20 pounds of force.

I developed loading concepts, wrote the evaluation protocol, recruited participants, and conducted a formative study to assess satisfaction, confidence, and whether the tube was pierced.

Control and three sample-loading concepts used in a usability evaluation
Control and three loading concepts · Formative evaluation
What the evaluation showed
6 / 10

Hospital staff preferred concept V3.

Ten trained hospital staff with varied roles participated. Six preferred V3, one preferred V1, and three had no preference or an unknown preference. The findings informed the loading direction and an instructional layer in the interface.

03 · Industrial design

Bring the experience together.

Alongside the interface, I helped define the device’s visual language and created a detailed surface model for engineering. The design used softened forms, texture, and breaks in large surfaces to reduce perceived size.

Designer Chet Larrow collaborated on enclosure ideation and exploration, refinement, color, material, and finish, and usability evaluation.

Quantra enclosure design development
Industrial design exploration and refinement.
The takeaway

Clarity is a system, not a screen.

A clinician’s experience spans a physical task, the information on a display, and the confidence to act. This project brought those pieces together through observation, prototyping, and formative evaluation.

Design the interaction. Test the assumption. Make the next decision clearer.
Andre Minoli · Product design and usability
Project work developed with Key Tech and Hemosonics · Visuals from the original case study